From: Design and testing of a 96-channel neural interface module for the Networked Neuroprosthesis system
No. Channels | Power Mode | System Power |
---|---|---|
1 | 30ksps | 31.2 mW |
1 | 2ksps | 22.1 mW |
96 | 2ksps DMA sampling + CPU awake | 45.3 mW |
96 | 2ksps DMA sampling + Disabled Peripherals + CPU awake | 40.0 mW |