Skip to main content

Table 2 Power Saving Techniques

From: Design and testing of a 96-channel neural interface module for the Networked Neuroprosthesis system

No. Channels Power Mode System Power
1 30ksps 31.2 mW
1 2ksps 22.1 mW
96 2ksps DMA sampling + CPU awake 45.3 mW
96 2ksps DMA sampling + Disabled Peripherals + CPU awake 40.0 mW
96 2ksps DMA sampling + Disabled Peripherals + CPU asleep 33.6 mW