From: Effect of uniform capacitively coupled electric fields on matrix metabolism of osteoarthritic cartilage

Fig. 1

Experimental Apparatus. A A CAD rendering of the fabricated in-vitro exposure apparatus consisting of a cell culture dish and a custom lid with integrated glass coverslips to serve as the dielectric for capacitive stimulation. B Cross-section of the FEA model of the electric field generated within the in-vitro exposure apparatus. The field uniformity is within 5% of the desired value of 20 mV/cm in all regions within the dish. C A stack of manufactured culture dishes within an incubator showing the attached wires used to deliver the signal