Skip to main content

Table 4 Event-wise interrater agreement of seizure annotations by modality

From: Using a standalone ear-EEG device for focal-onset seizure detection

Modality

Intracranial EEG

Scalp EEG

Ear-EEG

Events (#)

39

14

38

Events agreed (#)

22

14

26

FEA (%)

56.4

100

68.4

Events duration (s)

4041

1734

5022

Events agreed duration (s)

2939

1475

3120

FEDA (%)

72.7

85.1

62.1

  1. FEA fraction of event agreement, FEDA fraction of event duration agreement